Abdomen
289
Hints
• You could repeat the same protocol simply by click-
ing the chronicle with the right mouse button for
“repeat“. E.g. when both non-contrast and contrast
studies are required.
• Water, rather than positive oral contrast agents
should be used. Give the last cup 200 ml just prior to
positioning the patient. To ensure adequate filling of
the duodenal loop, lay the patient on the right side
for 5 minutes before performing the topogram.
• A pre-contrast examination is usually performed
only if no CT scans were previously acquired, to
exclude calculi in the common bile duct and to visu-
alize possible lesions in the liver.
Emotion 6 AbdSeq
kV 130
Effective mAs/
Quality ref. mAs
120
Rotation time 0.8 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 4.0 mm
Feed/Scan 12.0 mm
Kernel B41s
CTDI
Vol
13.08 mGy
Effective dose Male: 6.77 mSv
Female: 11.34 mSv
Contrast medium IV injection
Start delay 50 – 60 sec.
Flow rate 3.0 ml/sec.
Total amount 100 ml

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM

Related product manuals