Page #325 background imageLoading...
Page #325 background image
Pelvis
325
Emotion 16 Hip 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
100
Rotation Time 1.0 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 5.0 mm 0.7 mm
Kernel B70s B70s
CTDI
Vol
12.50 mGy
Effective Dose Male: 2.31 mSv
Female: 5.93 mSv
Emotion 6 Hip 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
100
Rotation Time 1.0 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 5.0 mm 1.25 mm
Feed/Rotation 9.0 mm
Pitch Factor 1.50
Increment 5.0 mm 0.8 mm
Kernel B70s B70s
CTDI
Vol
12.70 mGy
Effective Dose Male: 3.60 mSv
Female: 3.16 mSv

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV