Upper Extremities
337
Emotion 16 WristHR 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
40
Rotation time 1.0 sec.
Acquisition 4 x 0.6 mm
Slice collimation 0.6 mm
Slice width 2.0 mm 0.6 mm
Feed/Rotation 1.8 mm
Pitch Factor 0.75
Increment 2.0 mm 0.4 mm
Kernel U90s U90s
CTDI
Vol
5.52 mGy
Effective dose Male: < 0 mSv
Female: < 0 mSv
Emotion 6 WristHR 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
40
Rotation time 1.0 sec.
Acquisition 6 x 0.5 mm
Slice collimation 0.5 mm
Slice width 2.0 mm 1.0 mm
Feed/Rotation 2.6 mm
Pitch Factor 0.85
Increment 2.0 mm 0.6 mm
Kernel U90s U90s
CTDI
Vol
5.12 mGy
Effective dose Male: 0.01mSv
Female: < 0 mSv