Hint
For image reconstruction of soft tissue, use kernel
B30s/B31s.
Emotion 6 ExtrCombi 2
nd
reconstr.
kV 130
Effective mAs/
Quality ref. mAs
50
Rotation time 1.0 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 4.0 mm 1.25 mm
Feed/Rotation 9.0 mm
Pitch Factor 1.50
Increment 4.0 mm 0.8 mm
Kernel B41s B70s
CTDI
Vol
6.35 mGy
Effective dose Male: 0.01 mSv
Female: 0.01 mSv