Vascular
395
Emotion 16 Embolism 2
nd
reconstr.
kV 110
Effective mAs/
Quality ref. mAs
80
Rotation Time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 3.0 mm 0.75 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 3.0 mm 0.5 mm
Kernel B31s B20s
CTDI
Vol
6.40 mGy
Effective dose Male: 3.28 mSv
Female: 4.26 mSv
Emotion 6 Embolism 2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
80
Rotation Time 0.8/0.6 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 3.0 mm 1.25 mm
Feed/Rotation 10.8/9.0 mm
Pitch Factor 1.80/1.50
Increment 3.0 mm 0.8 mm
Kernel B31s B20s
CTDI
Vol
6.64 mGy
Effective dose Male: 3.25/3.21 mSv
Female: 4.23/4.18 mSv