Page #411 background imageLoading...
Page #411 background image
Vascular
411
Emotion 16 AngioRunOff 2
nd
reconstr.
kV 110
Effective mAs/
Quality ref. mAs
90
Rotation Time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 6.0 mm 1.5 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 6.0 mm 1.0 mm
Kernel B31s B31s
CTDI
Vol
6.48 mGy
Effective dose Male: 4.98 mSv
Female: 5.08 mSv
Emotion 6 AngioRunOff 2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
90
Rotation Time 0.8/0.6 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 6.0 mm 2.5 mm
Feed/Rotation 21.6 mm
Pitch Factor 1.80
Increment 6.0 mm 1.5 mm
Kernel B31s B31s
CTDI
Vol
6.39 mGy
Effective dose Male: 2.95 mSv
Female: 1.94 mSv