EasyManuals Logo

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #519 background imageLoading...
Page #519 background image
Children
519
* The conversion factor for a 7-year-old child, and a
scan range of 120 mm was used.
Emotion 16 Head
kV 110
Effective mAs/
Quality ref. mAs
230
Rotation time 1.5 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 4.0 mm
Feed/Rotation 10.6 mm
Pitch Factor 0.55
Increment 4.0 mm
Kernel C30s
CTDI
Vol
36.80 mGy
Effective dose Male: 2.83 mSv*
Female: 2.89 mSv*
Emotion 6 Head
kV 110
Effective mAs/
Quality ref. mAs
230
Rotation time 1.5 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 5.0 mm
Feed/Rotation 2.4 mm
Pitch Factor 0.40
Increment 5.0 mm
Kernel C30s
CTDI
Vol
42.32 mGy
Effective dose Male: 1.70 mSv*
Female: 1.69mSv*