Dose Information
55
Outside the topogram range, CARE Dose 4D will con-
tinue the scan with the last available topogram infor-
mation. Without a topogram, CARE Dose 4D cannot be
switched on. Repositioning of the patient on the table
and excessive motion of the patient must be avoided
between the topogram and the scan. If two topograms
of the same projection exist for one scan range, the
last acquired will be used for determining the (eff.)
mAs. If a lateral and a.p. topogram exist for one scan
range, both will be used for determining the (eff.)
mAs.