Children
583
* The conversion factor for a 7-year-old child, and a
scan range of 200 mm was used.
Emotion 16 BodyAngio 2
nd
reconstr.
kV 80
Effective mAs/
Quality ref. mAs
80
Rotation time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 3.0 mm 0.7 mm
Kernel B41s B31s
CTDI
Vol 2.56 mGy
Effective dose Male: 1.34 mSv*
Female: 1.66 mSv*
Emotion 6 BodyAngio 2
nd
reconstr.
kV 80
Effective mAs/
Quality ref. mAs
80
Rotation time 0.8/0.6 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0 mm
Slice width 3.0 mm 1.25 mm
Feed/Rotation 9.0 mm
Pitch Factor 1.50
Increment 3.0 mm 0.8 mm
Kernel B41s B31s
CTDI
Vol 2.64 mGy
Effective dose Male: 0.04/2.18 mSv*
Female: 2.50 mSv*