EasyManuals Logo

Siemens SOMATOM Manual&Nbsp;

Siemens SOMATOM
590 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #588 background imageLoading...
Page #588 background image
588
Children
* The conversion factor for a 8-week-old child, and a
scan range of 150 mm was used.
Emotion 16 NeonateBody
kV 80
Effective mAs/
Quality ref. mAs
35
Rotation time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2 mm
Slice width 5.0 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 5.0 mm
Kernel B41s
CTDI
Vol 1.02 mGy
Effective dose Male: 0.96 mSv*
Female: 1.07 mSv*
Emotion 6 NeonateBody
kV 80
Effective mAs/
Quality ref. mAs
35
Rotation time 0.8/0.6 sec.
Acquisition 6 x 2.0 mm
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 1.50
Increment 5.0 mm
Kernel B41s
CTDI
Vol 1.02 mGy
Effective dose Male: 3.23 mSv*
Female: 4.12 mSv*

Table of Contents

Other manuals for Siemens SOMATOM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SOMATOM and is the answer not in the manual?

Siemens SOMATOM Specifications

General IconGeneral
CategoryCT Scanner
Rotation Time0.25 seconds
Tube Voltage70-150 kV
Bore Size78 cm
Max Table Load250 kg
Gantry Aperture78 cm
Table Weight Capacity250 kg
X-ray Tube Voltage70-150 kV
Scan Time per Rotation0.25 seconds
SoftwareSyngo
ConnectivityDICOM

Related product manuals