Workflow Information
73
Additional Information
As soon as you define a new recon range, all recon
ranges will be shown in the topo segment. The two
numbers on the right-hand side at the beginning of
each recon range indicate the recon job the range
belongs to. The first number stands for the scan range,
the second number stands for the recon job to which
the range belongs. If no recon job is pending, only the
scan ranges are shown in the topo segment. Only one
number on the right-hand side at the beginning of
each scan range indicates which scan the range
belongs to.
If the first recon job is saved as an Oblique recon job,
RTD images are displayed after scanning and the
Examination task card is automatically switched to
3D reconstruction
Patient Browser:
for each double oblique recon job, one series is
added in the Patient Browser.
•If Auto Reference Lines is selected the correspond-
ing reference image is added to the 3D recon series.
All reconstructions are performed in the background
Do not use high resolution images
Do not use extended FoV
If no entry is selected in the chronicle, all open
reconstructions are automatically reconstructed.
•If Autorecon is selected on the Recon tab card, this
recon job (axial and oblique) will be automatically
reconstructed after scanning.