EasyManua.ls Logo

SKF IMx-1 - Sensors

SKF IMx-1
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INTEGRATION WITH SKF @PTITUDE OBSERVER
Enlight Collect IMx-1 System View
30 (101)
SKF Enlight Collect IMx-1 System
User Manual
Revision F
Connection:
Connected
Closed
Firmware Version: The version of firmware, installed in the gateway.
Sensors: The number of sensors linked to the gateway.
IP Address: The IP address that is used by the gateway.
2.3.2 Sensors
Similarly, for sensors the following properties are shown in the sensor table:
Name: Name given to the sensor.
Hardware ID: The unique identifier for the sensor, if the sensor is commissioned and
has successfully connected via the gateway to @ptitude Observer, it will show a
MAC address in this field.
Location: For a measurement sensor, Leaf or Mesh, the hierarchical location where
that sensor and measurement points have been created. For a relay node, this field
contains the descriptive location information entered by the user,
Figure 25.
Self diagnostics status: Displays the latest reported self-diagnostic status or in
case of multiple errors the item considered most critical. The status field will initially
be empty until the actual sensor status has been received.
"Ok" or an error of any of the following types. Values in brackets are decimal
values corresponding to the bit set in the mesh network information log, Self-
Diagnostic entry, when the error is true:
o Battery level low (1)
o External Flash memory failure (16)
o Configuration CRC failure (64)
o Firmware update error (128)
o (Watchdog reset (256)only available in the log file)
o Network instability (512)
Self-diagnostic errors may also raise a system or critical system alarm in @ptitude
Observer. Note that the network instability error can have significant ramifications for
sensor battery life and urgent action should be taken, refer to
IMx-1 sensor
troubleshooting
.
Double clicking a sensor entry or selecting it and then choosing Edit, will launch the
sensor properties dialog where extended status information can be viewed on the
second tab, named Status.