Press the Total key to finish the totaling transaction.
The display now shows the gross, tare, net, and unit count totals, along with the item
counter and the sequential number of the last transaction.
Available choices for displaying the total
Deleting the totaling memory
The following functions can be assigned to the Total function key:
•
Display, print, and delete total
After pressing the Total function key, the totals are displayed for 3 seconds, and the
memory contents are printed and deleted at the same time. A new totaling transaction can
then be started.
•
After pressing the Total function key, the user can decide whether to delete the totaling
memory.
Press the F2 Delete Total key to delete the totaling memory.
Press the F6 Back key to continue totaling without deleting the memory.
Relief factor when totaling
(cannot be enabled for picking)
You can specify in setup mode that the scale must be relieved by a certain number of number
increments between placing individual items
If the scale is not relieved, the weight value is not saved to the totaling memory, and the error
message "Error 55 - Place Weight" is issued. This for instance prevents saving an item twice.
The Auto Totaling function can be enabled in setup mode.
In this case, only the first item must be saved with the Plus key, all other items are automatically
saved to the totaling memory after the scale is loaded and settles.
The Plus key remains hidden as long as auto totaling is enabled.
If auto totaling is terminated with cancel, error message or change of measurement location,
auto totaling must be restarted with the Plus key.
Info: Just like manual totaling, auto totaling can be defined as a condition that triggers a printout
(see Section 8 Printing).
Assigning and deleting the sequential number
Every entry (plus, manual, cancel) into the totaling memory is assigned a 4-digit sequential number.
The sequential number is not reset in the factory default condition.
Resetting the sequential number can be optionally programmed in setup mode:
• disabled (no reset = factory default)
• on Delete Total key
• on switch off
• function key "Delete Seq. Number"