EasyManuals Logo

Sonel MPI-536 User Manual

Sonel MPI-536
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
MPI-536 USER MANUAL
79
Additional information displayed by the meter
Measurement in progress.
U
B
>U
L
!
The touch voltage exceeds a preset U
L
threshold value.
No U
L-N
!
Lack of neutral lead that is necessary for I
Δ
n constant and
pulsed with direct current offset
The meter is ready for measurement.
U
L-N
voltage is incorrect for making a measurement.
U
L-PE
voltage is incorrect for making a measurement.
U
N-PE
voltage is incorrect for making a measurement.
Phase connected to N terminal instead of L terminal (for exam-
ple, exchange of L and N in the mains socket).
Maximum temperature of the meter is exceeded.
Network frequency is outside the range of 45...65 Hz.
PE conductor connected incorrectly.
Measurement error.
Before measurement, voltage at test terminals exceeds 500 V.
Voltage exceeded.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonel MPI-536 and is the answer not in the manual?

Sonel MPI-536 Specifications

General IconGeneral
BrandSonel
ModelMPI-536
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals