Legend™ T / RT
SO
R
V
AL
L
®
EASYset™
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&

Table of Contents

Questions and Answers

  • C
    Carrie ReyesSep 10, 2025
    What to do if Sorvall Legend RT displays error E-14?
    • K
      Karen PruittSep 12, 2025
      If your Sorvall Laboratory Equipment displays error code E-14, switch the instrument off and on again. If the error persists, contact KENDRO service.
  • J
    John WilkersonSep 9, 2025
    What to do if Sorvall Laboratory Equipment displays error E-01 or E-13?
    • W
      williamsmarySep 9, 2025
      If your Sorvall Laboratory Equipment displays error code E-01 or E-13, switch the instrument off and on again. If the error persists, contact KENDRO service.
  • S
    smithshellySep 7, 2025
    What to do if Sorvall Laboratory Equipment displays error E-34?
    • K
      Kristy StokesSep 8, 2025
      If your Sorvall Laboratory Equipment displays error code E-34, switch the instrument off and on again.
  • T
    Tiffany ColeAug 23, 2025
    How to open the lid of Sorvall Legend RT?
    • M
      Melissa JonesAug 23, 2025
      If the lid of your Sorvall Laboratory Equipment won't open, first ensure the main power is on and the instrument display is lit. If it is, press down on the middle of the front section of the lid and then press the "open lid" key. As a last resort, use the emergency lid release.
  • E
    emckenzieSep 5, 2025
    What causes error E-18 on Sorvall Legend RT Laboratory Equipment?
    • A
      Alyssa DanielsSep 6, 2025
      To fix error E-18 on your Sorvall Laboratory Equipment, verify that the bucket is approved for use with the inserted rotor.
  • A
    Aaron JonesSep 4, 2025
    What causes error E-21 on Sorvall Laboratory Equipment?
    • F
      fergusonjenniferSep 4, 2025
      To address error E-21 on your Sorvall Laboratory Equipment, first check if a certified rotor is inserted. If the error occurred after a brief power failure, switch the instrument off and on again using the main switch.
  • E
    ericgregorySep 3, 2025
    What causes error E-20 on Sorvall Laboratory Equipment?
    • K
      kfisherSep 3, 2025
      To resolve error E-20 on your Sorvall Laboratory Equipment, first check if a certified rotor is inserted. Then, ensure the rotor is completely fastened onto the motor shaft by releasing the rotor fastening until you can easily lift and depress it, and then re-tighten the fastening.
  • S
    Stephanie AvilaSep 2, 2025
    What to do if Sorvall Laboratory Equipment displays error E-33?
    • B
      Brian HamiltonSep 2, 2025
      If your Sorvall Laboratory Equipment displays error code E-33, turn the instrument off and pull the main plug. Inspect and clean the ventilation slots if necessary. After about 60 minutes, you can restart the instrument, ensuring you observe the maximum permissible environmental temperature.
  • E
    Erica CannonSep 1, 2025
    What to do if Sorvall Legend RT displays error E-32?
    • A
      amanda59Sep 1, 2025
      If your Sorvall Laboratory Equipment displays error code E-32, turn the instrument off and unplug the mains power cord. Inspect and clean the ventilation slots if necessary. After about 60 minutes, restart the instrument, ensuring you observe the maximum permissible environmental temperature.
  • M
    margaretwilliamsAug 31, 2025
    What to do if Sorvall Legend RT displays error E-31?
    • P
      Pamela CruzAug 31, 2025
      If your Sorvall Laboratory Equipment displays error code E-31, turn the instrument off and unplug the mains power cord. Inspect and clean the ventilation slots if necessary. After approximately 60 minutes, you can restart the instrument.

Summary

For Your Safety

Improper Use

Details incorrect handling that can cause harm or damage.

Centrifuging Hazardous Materials

Safety guidelines for handling dangerous substances.

The Legend T; RT

Safety Systems

Details the built-in safety features designed to protect users and the instrument.

Function and Features

Comprehensive list of the centrifuge's operational capabilities and advanced functions.

Before Use

Centrifuge Transport and Installation

Procedures and precautions for moving and setting up the centrifuge.

Proper Location

Criteria for selecting a suitable and safe operating environment.

Main Connection

Instructions for correctly connecting the centrifuge to the power supply.

Operation

Maintenance and Care

Troubleshooting

Emergency Lid Release

Procedure for manually opening the lid during power failure.

Error Code Identification

Guide to diagnosing and resolving instrument error codes.

Contacting Kendro Service

Information on how to get technical support from the manufacturer.

Technical Data

Sorvall Legend RT Specifications

General IconGeneral
TypeBenchtop centrifuge
Maximum Speed15, 000 rpm
Temperature Range-10°C to +40°C
Programs99
TimerUp to 99 hours
Power Requirements50/60 Hz
Acceleration Time≤ 20 seconds
Deceleration Time≤ 20 seconds
Noise Level≤ 60 dBA