Instruction Manual
Legend™ T / RT
SO
R
V
AL
L
®
EASYset™
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&

Table of Contents

Questions and Answers

  • D
    David VangAug 29, 2025
    What to do if Sorvall Laboratory Equipment displays error E-29?
    • A
      Ashley GonzalezAug 30, 2025
      If your Sorvall Laboratory Equipment displays error code E-29, switch the instrument off and on again using the main switch. Open the lid and check if the rotor can turn freely. If the problem persists, contact KENDRO service.
  • J
    Jeffrey KimAug 28, 2025
    What causes error E-40 on Sorvall Legend RT Laboratory Equipment?
    • C
      Christopher HuntAug 28, 2025
      To address the E-40 error on your Sorvall Laboratory Equipment, open the lid and check the load. Ensure that there isn't a significant imbalance due to a missing bucket, that the tubes or adapters fit properly (preventing the rotor from rubbing against the chamber or cover), and that there are no microplate carriers on the motor cover.
  • A
    Allison WilliamsAug 27, 2025
    What to do if Sorvall Legend RT displays message "Lid"?
    • J
      Joshua GoodAug 27, 2025
      If the message "Lid" is flashing in the display of your Sorvall Laboratory Equipment, and the lid was opened manually during operation, switch the instrument off and then on again to continue the run.
  • H
    Hannah Jenkins MDAug 27, 2025
    What to do if Sorvall Legend RT Laboratory Equipment displays message “Lift Lid“?
    • C
      cindydawsonAug 27, 2025
      If the message "Lift Lid“ appears in the display of your Sorvall Laboratory Equipment, avoid placing any objects on the centrifuge lid and try lifting the lid slightly.
  • M
    Madison RodriguezAug 25, 2025
    What to do if Sorvall Legend RT Laboratory Equipment displays message "bAl"?
    • J
      Jason MartinezAug 26, 2025
      If the message "bAl" appears on the display of your Sorvall Laboratory Equipment, open the instrument by pressing the "open lid" key. Then, check to ensure the rotor is properly loaded, and inspect for any broken tubes or rotor damage that may have triggered the imbalance switch. Also, verify that the trunnions of the swinging bucket rotor are properly lubricated.
  • C
    crystalchurchAug 25, 2025
    What to do if Sorvall Laboratory Equipment is too noisy?
    • S
      Susan MeltonAug 25, 2025
      If your Sorvall Laboratory Equipment is making excessive noise, immediately press the "stop" key, or unplug the mains power cord in an emergency. Wait for the centrifuge to stop completely. Then, check if the rotor is loaded correctly and look for any broken tubes or damage to the rotor or motor that might be causing the noise.

Summary

For Your Safety

Improper Use

Details incorrect handling that can cause harm or damage.

Centrifuging Hazardous Materials

Safety guidelines for handling dangerous substances.

The Legend T; RT

Safety Systems

Details the built-in safety features designed to protect users and the instrument.

Function and Features

Comprehensive list of the centrifuge's operational capabilities and advanced functions.

Before Use

Centrifuge Transport and Installation

Procedures and precautions for moving and setting up the centrifuge.

Proper Location

Criteria for selecting a suitable and safe operating environment.

Main Connection

Instructions for correctly connecting the centrifuge to the power supply.

Operation

Maintenance and Care

Troubleshooting

Emergency Lid Release

Procedure for manually opening the lid during power failure.

Error Code Identification

Guide to diagnosing and resolving instrument error codes.

Contacting Kendro Service

Information on how to get technical support from the manufacturer.

Technical Data

Sorvall Legend RT Specifications

General IconGeneral
TypeBenchtop centrifuge
Maximum Speed15, 000 rpm
Temperature Range-10°C to +40°C
Programs99
TimerUp to 99 hours
Power Requirements50/60 Hz
Acceleration Time≤ 20 seconds
Deceleration Time≤ 20 seconds
Noise Level≤ 60 dBA