Return To Table of Contents
Operation Guide
OPERATING THE OPTIONAL SCALE SYSTEM - ELECTRIC LITTER OPTION WITH CHAPERONE (NOT
AVAILABLE WITH PRIME X OPTION) (CONTINUED)
Note: Do not touch the stretcher while the scale system is weighing or zeroing. The patient must remain still while
the system is weighing. If the patient is moving, the system will try for 20 seconds to get a stable weight or zero value
before displaying the error message .
If there is a loose connection or a malfunctioning component, the display will show “Err”. Attempt the function again. If
the malfunction is still present, the display shows “Err” again. Call Stryker technical support at 800-327-0770.
To meet the accuracy claim as stated in the product specification on page 11, the patient surface must be in the flat
position (fowler and gatch down) and the stretcher cannot exceed 5 degrees of Trendelenburg/reverse Trendelenburg.
OPERATING THE CHAPERONE (STRETCHER EXIT) OPTION (NOT AVAILABLE WITH PRIME X OPTION)
To use the chaperone with zone control option:
1. Press the Zero button to reset the scale system.
Note: Before positioning the patient on the stretcher, the scale system must be zeroed for the chaperone function
to operate properly.
2. Position the patient on the stretcher and press the Arm/Disarm button to activate the chaperone function. The
“Zone 1” LED will turn on. The chaperone function with zone control automatically selects Zone 1.
3. To select Zone 2 instead, press the Arm/Disarm button twice within three seconds of each other. The “Zone 2”
LED will turn on.
To deactivate the chaperone function, press the Arm/Disarm button. The selected Zone light will turn off.
CHARGING THE OPTIONAL SCALE SYSTEM BATTERY PACK - ELECTRIC LITTER OPTION WITH
CHAPERONE (NOT AVAILABLE WITH PRIME X OPTION)
To avoid completely draining the battery pack and having the optional scale system shut down, charge the battery pack
whenever only one of the charge indicator bars on the display (2) is black as shown on page 44.
The battery pack charges whenever the power cord is plugged into a properly grounded, hospital grade power source.
When the unit is stationary, you should plug the power cord into a power source whenever possible.
The optional scale system - with chaperone option only requires one 10.8V Li-Ion battery pack (0058-134-000). When
fully discharged, the battery pack requires approximately 3 hours of charging time to recharge.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker 1115 and is the answer not in the manual?

Stryker 1115 Specifications

General IconGeneral
CategoryMedical Equipment
Model1115
Frame MaterialSteel
Weight Capacity500 lbs
Mattress TypeFoam
IV PoleYes

Related product manuals