Chapter Five Maintenance Procedures Litter
107
5.5 HEAD SECTION (CONTINUED)
DRIVE SCREW AND BALL NUT REPLACEMENT HEAD MOTOR
(FOWLER BALL SCREW KIT P/N 3000300707)
Required Tools:
1/4 Nut Driver 5/16 Nut Driver 7/16 Socket Wrench
NeedleNose Pliers 7/16 Open End Wrench (2)
B
A
C
D
E
Procedure: (This procedure will take approximately 90 minutes)
1. Unplug the power cord from the wall socket.
2. Remove the head motor (see page 103) for access to the screw and ball nut.
3. Remove the head motor drive isolator and CPR decoupler (see page 105).
4. Using needlenose pliers, remove the cotter pin holding the clevis pin to the drive screw wing (A).
5. Remove the clevis pin and slide the wing and tapered roller bearing off the drive screw.
6. Using a 5/16 nut driver underneath the litter, remove the screw holding the cam retention wire (B) to the
ball nut assembly.