Chapter Six Maintenance Procedures Siderails
121
6.1 HEAD AND FOOT END SIDERAIL (CONTINUED)
HEAD END SIDERAIL CABLE REPLACEMENT
Required Tools:
#1 or #2 Phillips Screwdriver Side Cutters
A
B
C
D
Procedure: (This procedure will take approximately 25 minutes)
1. Run the head section fully up.
2. Unplug the bed power cord from the wall socket.
3. Put the siderail in the down position.
4. Remove the siderail cover (see page 117).
5. Using a Phillips screwdriver, remove the two screws holding the rear siderail pivot arm cover to the pivot
arm. Remove the cover to expose the siderail cables.
6. Using side cutters, clip the cable tie (A) holding the cables together.
6. Disconnect cable (B) from the circuit board, cable (C) from the nurse call and cable (D) from the speaker.
NOTE
The speaker and nurse call are optional equipment and may not be in the siderail as shown.
7. Pull the cables through the siderail (toward the center of the bed).