EasyManua.ls Logo

Stryker 3000 - Power Cord Replacement; Litter Grease Points

Stryker 3000
249 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter Five Maintenance Procedures Litter
116
5.9 POWER CORD REPLACEMENT
(POWER CORD KIT P/N 3000300737)
Required Tools:
Heyco Pliers 1/4 Nut Driver
Side Cutters 5/16 Nut Driver
Procedure: (This procedure will take approximately 15 minutes)
1. See page for proper routing of the power cord.
2. Unplug the power cord from the wall receptacle.
3. Remove litter access panels 3 and 4 (see page 94).
4. Using Heyco pliers, remove the power cord strain relief connector.
5. Using side cutters, cut all cable ties harnessing the power cord to other cables.
6. Using a 1/4 nut driver, remove the two clamps holding the power cord to the litter.
7. Unplug the power cord from the litter junction power board.
NOTE
Notice the plug location for proper installation of the new power cord.
8. Using a 5/16 nut driver, remove the screw holding the power cord ground cable to the base.
NOTE
Notice the arrangement of the star washers so the new ground wire will be installed properly.
9. Reverse the above steps to install the new power cord.
5.10 LITTER GREASE POINTS
1. All litter pivot joint bushings Syntech synthetic grease.
2. All Gatch and Fowler pivot link bushings Syntech synthetic grease.
3. Gatch drive screw and Gatch pan slide surface Syntech synthetic grease.
4. Fowler drive screw Mobil 28 grease.