Chapter Four Maintenance Procedures Base
68
4.1 BASE COVER REMOVAL (CONTINUED)
LOWERING BELLOWS
Required Tools:
5/16 Nut Driver #2 Phillips Screwdriver
A
B
HEAD END
Procedure: (This procedure will take approximately 5 minutes)
1. Unplug the power cord from the wall socket. Remove litter access panel #3 (A) (see page 94).
2. Using a 5/16 nut driver, remove the 2 screws (B) holding the bellows to the litter top (as shown in the illustra-
tion).
3. Allow the bellows to drop down.
NOTE
The power and sensor coil cords are secured to the bellows bracket. Pull the cords and guide them through
the plastic grommets at the lift header crossbar to allow adequate cord length for proper lowering of the bel-
lows.
4. Reverse the above steps to reattach the bellows and litter access panel.
NOTE
The procedure for lowering the bellows is the same for both ends of the bed except, when lowering the foot
end bellows, litter access panel #1 must be removed (see page 93) instead of panel #3.