Chapter Five Maintenance Procedures Litter
96
5.3 MOTION INTERRUPT SWITCH REPLACEMENT
Required Tools:
5/16 Nut Driver 1/4 Nut Driver
Standard Screwdriver
Procedure: (This procedure will take approximately 30 minutes)
1. Unplug the power cord from the wall receptacle.
2. Remove the motion interrupt panel (see page 95).
3. Remove appropriate litter access panel(s) (see pages 93 & 94).
4. Disconnect the two wires at the switch being replaced.
NOTE
Notice the terminals being used on the switch so the wires will be attached properly to the new switch.
5. Squeeze the retention clips on the switch and push down on the switch to remove it from the support
bracket.
6. Reverse the above steps to install the new switch.
7. Activate the motion stop system to assure it is functioning properly: press and hold down the BED DOWN
key. As the bed lowers, lift up on each corner of the motion interrupt pan individually and assure the down-
ward motion stops each time. Release the pan and allow the downward motion to continue.