Chapter Two MPS Operations
31
2.7 WEIGH SYSTEM OPERATION GUIDE
WEIGH SYSTEM USAGE
OPERATING THE SCALE BEFORE PUTTING A NEW PATIENT IN BED
Prepare bed for patient stay (linens, pillows, etc.).
Press and release SCALE ON. The scale monitor will read:
WEIGHING
XXX.X LB
Press and hold ZERO. The scale monitor will read:
HOLD TO ZERO WT.
RELEASE TO ZERO
Release ZERO. The scale monitor will now read:
DO NOT TOUCH BED
0.0 LB
The bed is now ready for the patient.
NOTE
Do not zero the bed while a patient is in bed. If this should occur, remove the patient and zero the bed.
OPERATING THE SCALE IF PATIENT IS ALREADY IN BED
If it is necessary to add or remove special equipment (monitors, pumps, etc.) during the patients
stay, press and release SCALE ON to activate the weigh system. After the scale monitor reads
XXX.X LB, press and hold . The scale monitor will read:
HOLD TO START
RELEASE TO START
Release . The scale monitor will read:
DO NOT TOUCH BED
ADD/REMOVE EQUIP
Add or remove the equipment and press . The scale monitor will read:
RELEASE TO FIN.
Release . The scale monitor will read:
DO NOT TOUCH BED
XXX.X LB
The weight displayed will be that of the patient only.

Table of Contents