Chapter Five Maintenance Procedures Litter
109
5.6 CPR
CPR CABLE ADJUSTMENT/REPLACEMENT
Required Tools:
NeedleNose Pliers (2) 10 mm OpenEnd Wrenches
NOTE
The CPR emergency release allows quick access to the patient when the head section is raised. When han-
dle (A) is squeezed, the attached cable (B) pulls the CPR bracket (C) which pulls the motor coupler out of
the drive screw and causes the head section to drop. If this doesnt happen when the handle is squeezed,
cable adjustment or replacement may be required.
Cable Adjustment Procedure: (This procedure will take approximately 5 minutes)
1. If the motor coupler is not pulled far enough away from the motor drive screw when the handle (A) is
squeezed to allow the head section to drop, use two 10 mm openend wrenches to loosen the two jam
nuts (D) holding the cable in position (one nut is not shown). Turn the nuts to allow less thread exposure
in area 1 (as shown in the illustration). When adjusted properly, the CPR coupler (E) and CPR wing (F)
should be separated, as shown in view two of the illustration on page 111, when the handle is squeezed
fully. When the handle is released, the CPR coupler (E) and wing (F) should fit tightly, as shown in view
one of the illustration. Tighten the nuts when the proper adjustment is achieved.
Cable Replacement Procedure: (This procedure will take approximately 15 minutes)
1. Unplug the bed power cord from the wall socket. Using needlenose pliers, remove the tension spring (G)
from the CPR bracket (C).
2. Remove the cable to be replaced from bracket (C) and bracket (H).
3. Remove the cable from the cable retainers (J).
4. Remove the cable from the bracket on the handle.
5. Using two 10 mm openend wrenches, remove the two nuts (D) (one not shown) and remove the cable
from the handle.
6. Reverse steps 15 to install the new cable. Refer to the procedure above for cable adjustment.
CPR HANDLE REPLACEMENT (CPR HANDLE KIT P/N 3000300719)
Required Tools:
NeedleNose Pliers #3 Phillips Screwdriver (2) 10 mm OpenEnd Wrenches
Procedure: (This procedure will take approximately 20 minutes)
1. Unplug the bed power cord from the wall socket. Using needlenose pliers, remove the tension spring (G)
from the CPR bracket (C).
2. Remove the end of the cable at the handle from bracket (K).
3. Using two 10 mm openend wrenches, remove the two nuts (D) (one not shown) and remove the cable
from the handle area.
4. Using a Phillips screwdriver, remove the two screws (L) holding the handle assembly to the litter and re-
move the handle.
5. Reverse the above procedure to install the new handle. See above procedure for cable adjustment.