Chapter Six Maintenance Procedures Siderails
125
6.1 HEAD AND FOOT END SIDERAIL (CONTINUED)
HEAD END SIDERAIL MODULE REPLACEMENT (SEE PAGE 15 FOR LIST OF SIDERAIL
MODULE KITS)
Required Tools:
#1 or #2 Phillips Screwdriver
A
C
B
D
C
Procedure: (This procedure will take approximately 15 minutes)
1. Unplug the bed power cord from the wall socket.
2. Remove the siderail cover (see page 117).
3. Unplug the ribbon cable(s) (A) from the module(s) (B) to be removed. Make note of the proper cable loca-
tions.
4. Using a Phillips screwdriver, remove the screws (C) holding the module in place and remove the module.
NOTE
If only the switch board needs to be replaced, and not the entire module assembly, remove only the screws
(D) holding the switch board to the module.
5. Reverse the above steps for installation of the new module.
NOTE
On all siderail modules, the switch boards can be replaced separately or as a kit that includes the outer label,
the switchboard and the ribbon cable.