Chapter Four Maintenance Procedures Base
87
4.6 LIFT UNIT (CONTINUED)
LIFT MOTOR COUPLER REPLACEMENT (LIFT MOTOR COUPLER KIT P/N 3000200705)
Required Tools:
Phillips Screwdriver Bungee Cord (or equivalent) 5/16 Socket Wrench
5/16 Nut Driver 1/2 Socket w/extension
B
A
FOOT END BOTTOM VIEW
Procedure: (This procedure will take approximately 30 minutes)
1. Unplug the power cord from the wall socket. Remove upper lift cover (see page 67).
2. Remove lower lift cover (see page 67).
3. Using a 3/8 socket with extension, remove the four nuts and washers (A) holding the isolation plate (B)
to the lift housing and lower the lift motor and isolation plate assembly to allow access to the coupler. Take
care not to drop the isolation plate onto the base.