Chapter Six Maintenance Procedures Siderails
117
6.1 HEAD AND FOOT END SIDERAIL
HEAD AND FOOT SIDERAIL COVER REMOVAL
Required Tools:
#1 or #2 Phillips Screwdriver
A
B
B
C
C
D
C
Removal Procedure: (this procedure will take approximately 10 minutes)
1. Unplug the bed power cord from the wall socket.
2. Using a Phillips screwdriver, remove the 6 phillips screws (A) holding the outside cover (B) to the siderail.
CAUTION
There are two ribbon cables (C) connecting the outside cover to the head end siderail. Be careful not to pull
on them when removing the cover.
3. To remove the cover, lift the siderail latch handle and pull out and down on the cover to clear the handle.
4. Pinch the release tabs on the ribbon cables (C) at the circuit boards and remove the cables from the siderail.
Make note of the proper location for the cables.
5. Reverse the above steps to reattach the cover. Be sure the panel spacers (D) are in place before attaching
the cover.
CAUTION
Do not snag the ribbon cables when installing the siderail cover.
NOTE
Siderail cover removal for the foot end rails is the same procedure without the ribbon cables and switch com-
ponents.