Troubleshooting
Problem
Possible Cause Solution
System operates but actuators move
slowly:
Power is possibly disconnected and
the bed operates by battery.
1. Power cable might have come
out of the control panel; check it.
2. Check the connections of the
power cable.
3. Check the socket.
4. Recharge the battery, call the
service center.
Control box beeps while actuators
are raised and actuators will not
move in the downward direction:
Control box actuator position
reference has been lost.
1. Unplug the power cord and turn
off the battery.
2. Plug in the power cord and turn
on the battery.
3. Run each upper actuator to its
highest position.