4 SM-7100 REV 01 www.stryker.com
Summary of safety precautions
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WARNING
Only use the input voltage and frequency as rated on the product.
Do not clean, service, or perform maintenance while the product is in use.
Always power off and unplug the power cord before cleaning, servicing, or performing maintenance.
Always power off the product and unplug the power cord from the wall outlet when large spills occur near the circuit
boards, cables, and motors. Remove the patient from the product, clean up the fluid, and have service personnel
inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical product.
Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation.
Do not spray cleaners directly onto the battery, control boxes, actuators, cables, or other electric equipment.
Do not use abrasive powder, steel wool, or similar materials that may damage the product surface.
Do not use Virex
®
TB for product disinfecting.
Do not use acid-based chemicals or flammable chemicals, such as gasoline, diesel, or acetone for cleaning
purposes.
Do not directly spray or saturate the siderail control panel, patient control pendant, or nurse control pendant with
cleaners.
Do not use sharp objects to clean the siderail control panel.
Always make sure to line the hex shaft with the hex in the caster. Test the caster to identify the direction of the
brake. Damage may occur if you install the caster incorrectly.
CAUTION
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
To minimize the risk of any electromagnetic interference, the product design follows the standard IEC 60601-1-2.
To avoid problems, use the bed in accordance with the EMC/EMI requirements in the EMC section of this
operations manual.
Portable and mobile RF communications equipment can produce electromagnetic interference with SV1. To prevent
interference caused by portable and mobile RF communications equipment, maintain a minimum distance between
portable and mobile communications equipment and SV1 according to the EMC section of this operations manual.
Do not steam clean, pressure wash, ultrasonically clean, or immerse any part of the product in water. Exposure to
water may damage the internal electric parts. These methods of cleaning are not recommended and may void this
product’s warranty.
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning.
Some cleaning products are corrosive in nature and may cause damage to the product if you use them improperly.
If you do not properly rinse and dry the product, you may leave a corrosive residue on the surface of the product
that could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void
your warranty.

Table of Contents

Other manuals for Stryker SV1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker SV1 and is the answer not in the manual?

Stryker SV1 Specifications

General IconGeneral
BrandStryker
ModelSV1
CategoryMedical Equipment
LanguageEnglish

Related product manuals