98 SK-7500 REV 00 www.stryker.com
EMC information
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
SV2 is intended for use in an electromagnetic environment specified below. The customer or the user of SV2 should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment
RF Emissions
CISPR 11
Group 1
SV2 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
SV2 is suitable for use in all
establishments other than domestic
and those directly connected to the
public low voltage power supply
network that supplies buildings used
for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Complies
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency communication services. The user might need to
take mitigation measures, such as relocating or re-orienting the equipment.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
SV2 is suitable for use in the electromagnetic environment specified below. The customer or the user of SV2 should
assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic fast
Transient/burst
IEC 61000-4-4
+2 kV for power supply
lines
+1 kV for input/output lines
+2 kV for power supply
lines
+1 kV for input/output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
+1 kV line(s) to line(s)
+2 kV line(s) to earth
+1 kV line(s) to line(s)
+2 kV line(s) to earth
Mains power quality is that
of a typical commercial
and/or hospital
environment.