www.stryker.com SK-7500 REV 00 11
Troubleshooting
Problem
Possible Cause Solution
Bed does not operate
1. Power cable does not plug to the
mains.
2. Power cable is damaged.
3. Nurse control unit or hand control
unit may be locked.
4. Control box does not operate
1. Ensure power cable and plug are
not damaged and then plug the
power cable to the mains.
2. Replace the power cable by the
authorized service personnel.
3. Nurse control Unit activation
button should be checked. (On-
off). Inspect the nurse control unit
or hand control unit and make
sure they are not locked.
4. Control box should be replaced,
call the technical service.
When the system operates, one of
the actuators does not operate and
control box gives a click sound
1. Actuator cable may be out of the
control box socket.
2. Actuator cable is damaged.
3. Actuator is damaged;
4. CPR wire remains stuck;
5. Control box is damaged.
1. Check socket connections of the
control box.
2. Actuator cable should be
replaced, call the service center.
3. Actuator should be replaced, call
the service center.
4. Inspect CPR wire and nut
adjustment.
5. Control box should be replaced,
call the service center.
When the system operates, one of
the actuators does not operate and
control box does not give a click
sound
1. Nurse control unit or hand control
unit may be partly locked.
2. Control unit is damaged,
3. Control box is damaged.
1. Inspect nurse control unit or hand
control and make sure they are
not locked.
2. Unplug side rail control unit from
the socket and check for function
with the nurse control unit. If
function is restored, their may be
a problem with one of the control
units. Call technical service.
3. Control box should be replaced,
call the service center.
Battery is possibly discharged and
does not give click sound
1. Battery is completely discharged
2. Battery is damaged or expired.
1. Recharge the battery.
2. Battery should be replaced, call
the service center.
System operates but actuators move
slowly
Power is possibly disconnected and
the bed operates by battery.
1. Power cable might have come
out of the control panel; check it.
2. Check the connections of the
power cable.
3. Check the socket.