68 SK-7500 REV 00 www.stryker.com
Service
Nurse control pendant replacement (Continued)
Procedure:
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Power down the bed and unplug the power cord from the wall outlet.
4. Using diagonal pliers, cut and discard the cable tie that secures the fowler (backrest) actuator cable to the wire
routing channel.
5. Remove the three cable routing covers that secure the gatch (legrest) actuator cable to the litter frame.
6. Unplug the nurse control pendant cable from the junction box.
7. Dispose of the nurse control pendant.
8. Reverse steps to reinstall.
9. Verify proper operation of the product before returning it to service.
Siderail control panel (inside and outside siderail) replacement
Tools required:
Flat screwdriver
Procedure:
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Unplug the power cord from the wall outlet.
4. Using a flat screwdriver, carefully push in and up in the slotted access hole to remove the control panel from the
siderail hoop.
5. Unplug the cable connecting the siderail control panel to the junction box.
6. Remove and discard the siderail control panel.
7. Reverse steps to reinstall.
8. Verify proper operation of the product before returning it to service.
Control box replacement
Tools required:
Diagonal pliers
Small flat screwdriver
3/8 in. drive ratchet
10 mm socket
10 mm combination wrench
Procedure:
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Power down the bed and unplug the power cord from the wall outlet.
4. Turn the battery off using the nurse control.
5. Raise and lock the lower leg section in the highest position.
6. Using a small flat screwdriver, un-clip the actuator cable retainer (A) from the control box (Figure 11 on page 69).
7. Remove and save the actuator cable retainer.
8. Using a small flat screwdriver, push in on the power cord lock and gently pull outward on the power cord.