70 SK-7500 REV 00 www.stryker.com
Service
Battery replacement (Continued)
8. Dispose of the battery according to local laws and procedures.
9. Reverse steps to reinstall.
10. Verify proper operation of the product before returning it to service.
Figure 12: Battery
Power cord replacement
Tools required:
#2 Phillips screwdriver
Procedure:
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Power down the bed and unplug the power cord from the wall outlet.
4. Remove the two screws (A) that secure the power cord connection cover (B) to the litter frame (Figure 13 on page
71).
5. Remove and discard the external power cord (C) from the internal power cord (Figure 13 on page 71).
6. Reverse steps to reinstall.
7. Verify proper operation of the product before returning it to service.