AMI Sample Sequencer
Program List and Explanations
9. Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
edged). If an active error is acknowledged, the alarm relay opens
again. Cleared errors are moved to the Message list.
1.2 Message List
1.2.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation: View the Designation of instrument.
Version: Firmware of instrument (e.g. V5.00-12/16)
Version: ROVALVE 1.50
2.1.4 Factory Test: Test date of the Instrument and Motherboard.
2.1.5 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 Miscellaneous:
2.2.1.1 Case Temp: Shows actual temperature in [°C] inside the transmit-
ter.
2.3 Signal Outputs
2.3.1 Signal Output 1 to 6: Review the actual measured value in mA.
2.4 Signal Inputs
2.4.1 Signal Input 1 and 2: Review the actual measured value in mA.