AMI Sample Sequencer
Error List
A-96.250.661 / 120418 53
7. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error (blinking symbol)
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (e.g. Sample Flow low).
Such Errors are marked E0xx (bold and orange)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx (bold and red)
Error or fatal Error
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and
take corrective action.
Press [ENTER].
Navigate to menu Messages.
Press [ENTER].
Navigate to menu Pending Errors.
Press [ENTER].
Press [ENTER] to acknowledge the
Pending Errors. The Error is reset and
saved in the Message List.
RUN
15:20:18
0.0 l/h
1
Installation
Diagnostics
Messages
Maintenance
Main Menu
1.1
Pending Errors
Message List
Messages
1.1.5
Pending Errors
Error Code E010
Sample Flow low
<Enter> to Acknowledge