20 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
2.3. Instrument Specification
Power supply Available as an AC or DC model.
Voltage: 100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption: max. 8 VA
Sample
requirements
Flow rate: according to the attached analyzer
plus min. 5 l/h per stream
Sample pressure inlet: 0.53 bar (243 PSI)
Sample temperature: 545 °C (41122 °F)
NOTICE: The sample must not contain any oil, grease, or
sand.
On-site
requirements
The analyzer site must permit connections to:
Sample inlet (6): Tube 4 x 6 mm
Sample outlet (1): Tube 4 x 6 mm
Sample overflow (3): each 15 x 20 mm (1/2” hose nozzle)
Outlet pressure: atmospheric.
Electronics
housing
Aluminum
with a protection degree of IP 66 / NEMA 4X
Ambient temperature: -10 to +50 °C
Limit range of operation: -25 to +65 °C
Storage and transport: -30 to +85 °C
Humidity.: 10 to 90 % relative, non condensing
Display: backlit LCD, 75 mm x 45 mm