60 A-96.250.661 / 120418
AMI Sample Sequencer
Program List and Explanations
5 Installation
5.1 Sequence
5.1.1 Mode: Internal, AMI
Mode Internal: Set Mode to Internal if the AMI Sample Sequencer
is connected to an AMI Sodium A or an AMI Sodium P. In this Mode
the AMI Sample Sequencer automatically switches from sample 1
to max. 6.
Mode AMI: Set Mode to AMI if the AMI Sample Sequencer is con-
nected to an AMI Silica, AMI Silitrace or AMI Phosphate HL. In this
Mode the AMI Sample Sequencer works as a slave of the AMI ana-
lyzer. No further settings are available in the Mode AMI.
NOTICE: Do not set the AMI Sample Sequencer to Mode
Internal if it is connected to an AMI Silica, AMI Silitrace or AMI
Phosphate HL.
5.1.2 Signal Inputs: The signal inputs are used to receive the measuring
signal(s) from the current output of the AMI Sodium A or an AMI
Sodium P. The signal(s) are then transferred to a remote device
where the measuring value(s) are displayed.
If one signal input is selected, six sample channels are avail-
able.
If two signal inputs are selected, three sample channels are
available.
Range: 1 or 2
5.1.3 Flush Time: Set the flush time duration after channel switching.
Range: 103600 sec
5.1.4 Valid after: Set the remaining measuring time for channel x after the
6-way valve has switched to the next channel.
Range: 03600 sec
5.1.5 Cycle Time
Set the cycle time of each Channel (1 to max. 6). If the cycle time is
set to 0, the channel is switched off.
Range: 0600 min