AMI Sample Sequencer
Program List and Explanations
A-96.250.661 / 120418 61
5.2 Sample Flow
5.2.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E009 is displayed in the mes-
sage list.
Range: 0.00–50.00 l/h
5.2.2 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E010 is displayed in the message
list.
Range: 0.00–50.00 l/h
5.2.3 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluc-
tuates around the alarm value.
Range: 0.5–5 l/h
5.2.2 Delay: Duration, the activation of the alarm relay is retarded after
the messuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 sec
5.3 Miscellaneous
5.3.1 Language: Set the desired language.
Available settings: German/English/French/Spanish
5.3.2 Set defaults: Reset the instrument to factory default values in three
different ways:
calibration
in parts
completely
5.3.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
5.3.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the menus <Messages>, <Maintenance> and
<Installation>.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.