1.10 REPLACING THE LITHIUM MODULE
If the following message appears on the screen when the unit is switched on :
This means that the data back-up lithium module is flat (average life 10 years)
Note : It is possible to use D100S unit with a “flat” lithium module, but all
lithium hors service, seulement tous parameters and functions entered by the operator will be lost each time the
unit is switched off.
The module is replaced as follows :
After obtaining a new lithium module from Sylvac representative,
Place the unit on a table and remove the 4 retaining screws of the cover..
Touch a water pipe or other object connected to ground to release any static build-up (the inside of the unit is
Sensitive to electrostatic discharges).
Remove yellow cover.
Remove lithium module with a screwdriver :
Insert new lithium module, replace cover and replace the 4 cover retaining screws.
When the unit is switched on again, the memory contents lost message will appear once more.