4.5 TECHNICAL SPECIFICATIONS
Accuracy using extension cables :
These measuring errors are applicable only when using D100S unit without re-calibration :
Normal cable + extension up to 5 m: additional error 1.5 µm approx.
+ extension up to 10 m : 3 µm approx.
+ extension up to 15 m : 6 µm approx.
+ extension up to 20 m : 8 µm approx.
Direct cable + extension up to 5 m : additional error 3 µm approx.
+ extension up to 10 m : 6 µm approx.
+ extension up to 15 m : 10 µm approx.
This in a progressive error margin and re-calibration of D100S unit can considerably reduce the error.