Diagnostics
6–42 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 63
no “B” position
Problem: A GPIB command was sent to the
Autosampler CE to load or unload
sample B from a DSC cell.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 64
no pan in ref
(no pan in reference position)
Problem: The Autosampler CE has been
commanded to unload a pan from an
empty cell reference position. Re-
member that the Autosampler CE
always assumes that the DSC cell is
empty whenever you press one of the
Reset Sampler keys.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC