Instrument Control, and retry your
operation. If there is a pan in the
reference position, remove it manually
with tweezers.
GPIB Error 65
no pan in sample
Problem: The Autosampler CE has been
commanded to unload a pan from an
empty cell sample position. Remember
that the Autosampler CE always
assumes that the DSC cell is empty
whenever you press one of the Reset
Sampler keys.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation. If there is a pan
the sample position, remove it manu-
ally with tweezers.
GPIB Error 68
rec buf overflow
(receive buffer overflow)
Problem: The Autosampler CE GPIB com-
mand-receiving buffer has overflowed.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.