RSA-G2 Getting Started Guide Page 34
Choosing a Location
It is important to choose a location for the instrument using the following guidelines. The RSA-G2 should
be:
In
A temperature- and humidity-controlled area. Temperatures should be in range of 20 to 30°C.
A clean, vibration-free environment, preferably on the ground floor in the building. It should be
located away from pumps, motors, or other devices which produce vibrations.
An area with ample working and ventilation space.
An area that allows at least six inches of clearance on all sides of the equipment, and that allows a min-
imum of eight inches from the open and closed positions of all heated environment control systems.
An area that does not restrict airflow to cooling fans on the rear of the power control box, or the option
controllers.
On
A stable work surface. A marble table is recommended. Isolation mounts on a standard lab bench are
not recommended.
Near
A power outlet. An RSA-G2 with an oven requires 180 to 264 VAC/50 or 60 Hz with a 20 amp outlet.
Your TA Instruments TRIOS Instrument Control Software.
Compressed lab air and purge gas supplies with suitable regulators and filters.
Away from
Dusty environments.
Exposure to direct sunlight.
Direct air drafts (fans, room air ducts).
Poorly ventilated areas.
Noisy or mechanical vibrations.
High traffic areas, where constant movements from passing personnel could create air currents or
mechanical disturbances.