59
7.9.4.3
Retransmit Last Message
Send (0xC0 + Address) (1 byte binary
-
send 2 times sequentially). The suppl
y will return
the last message sent. Note that the supply does not have to be the active addressed su
p-
ply.
This command will not execute if another command is being processed.
Note that Single byte commands do not load data into the supply’s data output bu
ffer. Thus
this command will not cause the supply to retransmit data obtained from any previous Si
n-
gle Byte Co
m
mand.
7.9.4.4
Test if MD Mode is Installed
Send AA Hex followed by the address of the supply in binary. If not installed, the supply will
return a ‘1’.
If installed, the supply will return a ‘0’.
7.9.5
Addressed commands without response
7.9.5.1
Acknowledge SRQ
Send (0xE0 + Address) (1 byte binary
-
send 2 times sequentially). The supply will stop re
-
sending SRQ. If Enable SRQ retransmission is active, it will remain a
ctive.
7.9.5.2
Re
-
enable SRQ with out reading/clearing the SEVE Register
Send A5 Hex followed by the address of the supply in binary and new SRQ’s generated by
new events in the Fault Event will be enabled without reading and clearing the Status Event
Register. Al
l events previously recorded in the Fault Event Register must have been se
r-
viced by the user’s software prior to this command to take affect.
Name
Bit Positions
Response
Description
Global
Commands
Disable MD Mode
1010 0000
None
Set suppl
ies out of MD Mode (d
e-
fault)
Enable MD Mode
1010 0001
None
Set supplies into MD Mode
Disable SRQ
r
e
transmission
1010 0010
None
Disable retransmission of SRQs by
supplies (default)
Enable SRQ
r
e
transmission
1010 0011
None
Enable retransmission of SRQ
s by
supplies
Enable FLT Bit
1010 0100
None
Enable the FLT bit in the SENA
Re
g
ister
Disconnect serial
communications
1011 1111
OK
All supplies will halt transmission and
enter the non
-
addressed state.
Addressed Co
m-
mands
Read Registers
100x xxxx
R
egister data
Non destructive read of all register.
x xxxx is the address of the supply in
binary.