96
7.12.3 Voltage Subsystem
VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] <NRf+> Set output voltage
:TRIGger <NRf+> Set voltage value for trigger pending
:PROTection
[:OVER] <NRf+> Set OVP value
: LOW
:STATe <CRD> Set UVP|UVL mode
:[LEVel] <NRF+> Set UVP|UVL value
:CLEar
:MODE <CRD> Select arbitrary trigger control mode
VOLTage
Sets the output voltage value in Volts. The range of voltage values are described in Table 7-3. The
maximum number of characters is 12.
SCPI Command Syntax [SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
GEN Command Syntax PV < NRf+> PV?
Default Suffix V
*RST Value 0
Examples :VOLT 500 MV VOLT:LEV 234.56789
Query Syntax
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
VOLTage? MAX VOLTage? MIN
Returned Parameters
<NR3> VOLT? returns the present programmed voltage level.
VOLT? MAX and VOLT? MIN return the maximum and minimum programmable
current levels.
VOLT:MODE
This command selects FIX, LIST, WAVE subsystems control over the power supply output voltage.
SCPI Command Syntax [SOURce]:VOLTage:MODE <CRD>
GEN Command Syntax None
Parameters NONE | FIXed | LIST | WAVE
*RST Value NONE
Examples VOLT:MODE LIST VOLT:MODE FIX
Query Syntax [SOURce]:VOLTage:MODE?
Returned Parameters NONE | FIX | LIST | WAVE