Specification-2445A/2455A Service
Table
1-7 (cont)
Characteristics
Performance Requirements
Bench Handling (cabinet on and cabinet off) MIL-STD-810C, Method 516.2, Procedure V (MIL-T-28800C, para.
4.5.5.4.3).
Topple (operating with cabinet installed)
Set on rear feet and allow
to
topple over onto each
of
four adjacent
faces (Tektronix Standard 062-2858-00).
Packaged Transportation Drop
Meets the limits
of
the National Safe Transit Assn., test procedure
1 A-B-2; 10 drops
of
36 inches (Tektronix Standard 062-2858-00).
Packaged Transportation (Vibration)
Meets the limits
of
the National Safe Transit Assn., test procedure
1 A-B-1 ; excursion
of
1 inch p-p at 4.63 Hz
(1.1
g) for
30
minutes
(Tektronix Standard 062-2858-00).
EMI (Electro-magnetic Interference)
Meets MIL-T-28800C; MIL-STD-461 B, part 4 (CE-03 and CS-02),
part 5 (CS-06 and RS-02), and part 7 (CS-01, RE-02, and RS-
03)-limited
to 1 GHz; VDE 0871, Category B; Part 15 of FCC
Rules and Regulations, Subpart
J,
Class A; and Tektronix
Standard 062-2866-00.
Electrostatic Discharge Susceptibility Meets Tektronix Standard 062-2862-00. The instrument will not
change control states with discharges
of
less than 10 kV.
X-Ray Radiation
Meets requirements
of
Tektronix Standard 062-1860-00.
1-20