Performance Check-2445A/2455A Service
j. Move the probe to the
Channel
2 input connector.
k.
Set the VERTICAL MODE buttons to deselect Chan-
nel
1
and
display Channel 2.
NOTE
Instruments with
TV
OPTION 05 have a TV CLAMP
feature that
is
enabled by pushing
the
upper
CH
2
INPUT COUPLING button while
in
AC COUPLING.
The
letters
TVC
appear
in
the top right readout
when
this mode
is
selected. Push the lower
CH
2
INPUT COUPLING button to return to normal
AC
coupling.
I.
Repeat parts c through i for Channel 2.
m.
Disconnect the test setup.
3. Check
CH
1 and
CH
2 VOLTS/DIV,
CH
2
INVERT,
þÿ”V
and TRIGGER LEVEL Readout
Accuracies, Variable VOLTS/DIV, Vertical
Linearity, and
ADD.
a.
Set:
NOTE
Select channels to set
VOL
TS/DIV.
VOLTS/DIV
CH
1
and
CH
2 2
mV
VERTICAL MODE
CH
1
CH
2
CH
1
and
CH
2
BW
LIMIT
HORIZONTAL
A SEC/DIV
þÿ”V
TRIGGER
MODE
COUPLING
4-6
On
Off
1 þÿM©
DC
On
1 ms {knob
in)
On
{press
and
release for
a
ΔV
V readout)
AUTO
DC
NOTE
The
instrument must have had
at
least 20
minutes warmup prior to performing the following
steps.
b.
Momentarily press
and
hold
both the
CH
1 and
CH
2
upper Input Coupling buttons until a moving dot display
replaces the normal signal. This performs a
DC
Balance of
CH
1
and
CH
2 and the readout displays
DC
BALANCE
IN
PROGRESS.
c.
When
the signal
and
readout displays automatically
return to normal, set the
CH
1
and
CH
2 Input Coupling to
1 þÿM©
DC.
d.
Connect the Calibration Generator to the
CH
1
OR
X
input connector via a
50
þÿ©
BNC
cable.
Do
not use a
termination.
e.
CHECK-CH
1 and
CH
2 VOL TS/DIV, þÿ”V,and
TRIGGER LEVEL readout accuracies
as
follows:
1. Set VOL TS/DIV control to the first position listed
in
Table 4-2.
2. Set the Calibration Generator STD AMPLITUDE
output
level
to the corresponding Standard
Amplitude Input Level
in
Table 4-2.
NOTE
To
properly verify
TRIGGER
LEVEL
Readout Accuracy, the Calibration
Generator's STD AMPLITUDE output must
have rising and falling transition
times
(10% to 90%) > 20
ns.
No
overshoot
should appear on the waveform.
3. Verify that the generator output meets the
requirements noted above.
4.
Use
the VERTICAL POSITION control to set the
bottom of the signal
2 divisions below graticule
center.
5.
Rotate
þÿthe ”
REF
OR
DLY
POS
control to align
the reference cursor with the bottom of the
waveform.