3 Product-specific safety instructions
8
3 Product-specific safety instructions
DANGER
Integrated magnet
Danger to life for persons with pacemakers!
- Keep a distance of at least 20 cm between your pacemaker and the
measuring instrument.
ATTENTION
Integrated magnet
Damage to other devices!
- Keep a safe distance from devices that may be damaged by magnetism
(e.g. monitors, computers, credit cards, memory cards, etc.).
CAUTION
The condensate may be acidic.
Risk of burns to the hands!
- Wear acid-resistant safety gloves, glasses and overalls to empty the
condensate.
• Make sure that the condensate has been fully emptied out of the condensate
trap before the measuring instrument is stored for a long time.
• Before disposing of the product, the condensate trap must be emptied and
the condensate in the crude gas tube disposed of in a suitable container.
• When testing a gas pipe, pay attention to the following:
WARNING
Dangerous mixture of gases
Danger of explosion!
- Make sure there are no leaks between the sampling point and the measuring
instrument.
- Do not smoke or use naked flames during the measurement.