EasyManua.ls Logo

TESTO 300 - CO;NO Sensor Protect; NO2 Addition; O 2 Reference

TESTO 300
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13 Using the product
48
13.2.6.9 CO/NO sensor protect
Limit values can be set to protect the CO/NO sensors against overload. Sensor
protect is enabled if these are exceeded:
Fresh air dilution if exceeded (only for instruments with the "Dilution" option)
Shutdown if exceeded again
When dilution is enabled, the CO and CO undiluted values are
displayed in a blue font. A “*” is shown on the printout after the name of
both values to indicate dilution.
1
Call up function: | Device Settings | Sensor Protect
CO input screen: Set Sensor Setting is opened.
2
Enter the alarm limit value via the keypad.
3
Confirm input with .
The limit values must be set to 0 ppm to disable sensor protect.
13.2.6.10 NO2 addition
1
Call up function: | Device Settings | NO2 addition
2
Enter the addition value using the keyboard.
3
Confirm the input with OK.
4
Back to the Device Settings menu: tap .
13.2.6.11 O
2
reference
The O2 reference value of the current fuel can be set.
1
Call up function: | Device settings | O2 reference
O
2
reference current fuel input screen is opened.
2
Enter the value via the keypad.
3
Confirm input with .

Table of Contents

Other manuals for TESTO 300

Related product manuals