14 Performing the measurement
75
14.2.10 O
2
supply air
An O
2
dual wall clearance probe (0632 1260) must be connected.
Call up function: | O
2
supply air
The function is only available if the chosen fuel is a gas.
The gas burner capacity is calculated by means of the gas amount consumed.
To this end, a gas amount is input and its consumption read out at the gas
meter.
Call up function: | Gas Flow
Set the gas amount to watch for at the gas meter.
Set the calorific value of the burnt gas.
The test time is displayed.
When the set gas amount is reached: tap .
The calculated gas flow and the gas burner capacity (in KW) are
displayed.